Issue 06/2020 Art. No. 7001-0365
Operating Manual
Translation of the original operating manual
MKF (E5)
Alternating climate chambers with program control
Model Model version
Art. No.
MKF 56 MKF056-230V 9020-0378, 9120-0378
MKF056-240V 9020-0389, 9120-0389
MKF 115
MKF115-400V
9020-0379, 9120-0379
MKF115-400V-C 9020-0357 (with voltage and frequency changer)
MKF 240
MKF240-400V
9020-0380, 9120-0380
MKF240-400V-C
9020-0358 (with voltage and frequency changer)
MKF 720 MKF720-400V 9020-0381, 9120-0381
MKF720-400V-C
9020-0359 (with voltage and frequency changer)
MKFT (E5)
Alternating climate chambers with deep temperature
with program control
Model Model version
Art. No.
MKFT 115 MKFT115-400V 9020-0382, 9120-0382
MKFT115-400V-C
9020-0362 (with voltage and frequency changer)
MKFT 240 MKFT240-400V 9020-0383, 9120-0383
MKFT240-400V-C
9020-0361 (with voltage and frequency changer)
MKFT 720
MKFT720-400V
9020-0384, 9120-0384
MKFT720-400V-C 9020-0360 (with voltage and frequency changer)
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • J
    James VaughnSep 9, 2025
    What to do if my Binder MKF 115 chamber doesn’t heat up?
    • B
      Brittney WoodSep 9, 2025
      If your Binder Laboratory Equipment chamber isn’t heating up, it could be due to a defective heating element, in which case you should contact BINDER service.
  • S
    Sharon MedinaAug 23, 2025
    What to do for condensation or icing at the sides of the inner chamber of Binder Laboratory Equipment?
    • E
      Ellen MontoyaAug 23, 2025
      If you observe condensation or icing at the sides of the inner chamber, it's likely because the set-point has been below the ambient temperature for an extended period, causing icing in the preheating chamber. Defrost the chamber.
  • H
    hendersonsherryAug 25, 2025
    Why am I getting a humidity alarm message when operating without humidity on my Binder Laboratory Equipment?
    • A
      ashley10Aug 26, 2025
      If you receive a humidity alarm message while operating without humidity (humidity switch OFF), it's because the humidity set-point is set to a value greater than 0% r.H. In Manual Mode, enter a humidity set-point of 0% r.H. In Program Mode, enter a humidity subprogram with a humidity set-point of 0% r.H.
  • K
    Katherine MejiaAug 27, 2025
    Why is there no or low dehumidification in my Binder MKF 115?
    • M
      michael16Aug 27, 2025
      If there is no or low dehumidification, it could be because the capillary tube is blocked. Contact BINDER service.
  • T
    Terri BishopAug 27, 2025
    What causes icing at the sides of the inner chamber in Binder MKF 115?
    • B
      bradley05Aug 28, 2025
      If there's icing at the sides of the inner chamber, it's likely because the set-point has been below the ambient temperature for too long. Defrost the chamber.
  • B
    browntinaAug 29, 2025
    How to resolve humidity fluctuation in Binder MKF 115 Laboratory Equipment?
    • M
      Michael PerezAug 29, 2025
      If the control accuracy of ± 2.5 % r.H. is not reached, it may be due to a defective door gasket. Replace the door gasket.
  • U
    umooreSep 18, 2025
    What to do if there is no dehumidification and a "1H PREHEAT PHASE" notification in the controller display of my Binder MKF 115?
    • R
      Richard HolmesSep 18, 2025
      If there is no dehumidification and the notification "1H PREHEAT PHASE" is displayed, it means the rear power switch was turned on less than 1 hour before operating the chamber. Turn on the rear power switch at least one hour before operating the chamber.
  • K
    Kevin JacksonSep 18, 2025
    What to do if the Binder MKF 115 Laboratory Equipment chamber has no function?
    • K
      Kenneth GibsonSep 19, 2025
      If the Binder Laboratory Equipment chamber has no function and turning on the main power switch has no effect, it means that the rear power switch is not turned on. Turn on the rear power switch at least one hour before operating the chamber.
  • T
    Thomas StephensSep 7, 2025
    Why is there no refrigerating performance in my Binder Laboratory Equipment?
    • C
      clarkbrendaSep 8, 2025
      If there is no refrigerating performance, it may be because the ambient temperature is greater than 25 °C / 77°F. You should select a cooler place for installation.
  • T
    Troy VasquezSep 4, 2025
    What should I do if the safety device class 2 responds in my Binder Laboratory Equipment?
    • A
      Amber WilliamsSep 4, 2025
      If the safety device class 2 responds, it means the limit temperature has been reached. Disconnect the Binder Laboratory Equipment chamber from the power supply and allow it to cool down. Identify and eliminate the cause of the over-temperature. Press the 'RESET' button on the controller. Restart the chamber and verify the control functions. If necessary, choose a more appropriate limit value.