Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
- LLaura WaltersSep 23, 2025
The chamber might be fully loaded. Try loading the chamber with fewer items or consider allowing longer heating-up times.
Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
The chamber might be fully loaded. Try loading the chamber with fewer items or consider allowing longer heating-up times.
Why is the set-point temperature not reached in my Binder Laboratory Equipment after the specified time?
Several factors could be the reason. Make sure the chamber door is completely closed and that the door gasket isn't defective (replace if needed). The controller may also need calibration and adjustment. Finally, verify the power supply for the correct voltage.
What to do if Binder BD 115 Laboratory Equipment chamber doesn’t heat up but the controller display is working?
The timer might have run off. Try re-programming the timer or turning it off.
Why is my Binder Laboratory Equipment chamber heating permanently and not holding the set-point?
The controller may not be adjusted, so calibrate and adjust it. If the issue persists, the controller might be defective and further assistance is needed.
What to do if the Binder BD 115 Laboratory Equipment chamber has no function and the standby icon is displayed?
The chamber is likely in standby mode. Press down the standby button until the display lights up.
What to do if Binder Climate chamber fan is not working?
If the fan in your Binder Climate chamber isn’t turning or is turning too slowly, the fan might be defective, in that case contact BINDER service.
Why does my Binder Climate chamber take longer to heat up?
If you're experiencing deviations from the indicated heating-up times with your Binder Climate chamber, it might be because the chamber is fully loaded. Try loading the chamber with fewer items or allow for longer heating-up times.
Why is my Binder Climate chamber in stand-by mode?
If your Binder Climate chamber is without function and only the green 'stand-by' LED is lit, the chamber is likely in stand-by mode. Press the ON/OFF button until the display lights up.
Why is my Binder Climate chamber not working?
Your Binder Climate chamber might not be working due to several reasons. First, check the connection to the power supply. Second, the chamber fuse might have responded, so check it and replace it if needed. If it responds again, contact BINDER service. Finally, the controller could be defective, in which case you should also contact BINDER service.
What to do if my Binder BD 115 Laboratory Equipment chamber is heating permanently, and the set-point is not held?
The controller may not be adjusted correctly. Calibrate and adjust the controller.
Qualification required for installation, testing, and operation.
Explanation of safety symbols, signal words, and pictograms.
General safety guidelines for installation and operation.
Proper uses, improper uses, and potential misuses of the chamber.
Procedures for unpacking, checking delivery, and safe transport.
Requirements for installation site, storage, and environmental conditions.
Detailed information on connecting the chamber to the power supply.
Procedures for loading the chamber, including weight limits.
Explanation of controller buttons, display, and status symbols.
How the controller shows normal operating status and data.
Procedure to change the controller's display language.
How to set the internal clock and date on the controller.
Method to switch between Celsius and Fahrenheit units.
Entering target temperature and fan speed values via the controller.
Controlling air exchange by adjusting the position of the air flap.
Setting or changing access passwords for controller functions.
Function of the class 1 overtemperature safety device.
Overview of safety controller classes and function checks.
Procedure for setting the safety controller limit temperature.
How to interpret and respond to alarm messages.
Overview of available timer functions: Delayed Off, Temp. Dep. Delayed Off, Delayed On.
Setting up and activating the "Delayed Off" timer function.
How to stop or modify the "Delayed Off" timer settings.
Common faults and their solutions for operating personnel.
Qualification required for installation, testing, and operation.
Explanation of safety symbols, signal words, and pictograms.
General safety guidelines for installation and operation.
Proper uses, improper uses, and potential misuses of the chamber.
Procedures for unpacking, checking delivery, and safe transport.
Requirements for installation site, storage, and environmental conditions.
Detailed information on connecting the chamber to the power supply.
Procedures for loading the chamber, including weight limits.
Explanation of controller buttons, display, and status symbols.
How the controller shows normal operating status and data.
Procedure to change the controller's display language.
How to set the internal clock and date on the controller.
Method to switch between Celsius and Fahrenheit units.
Entering target temperature and fan speed values via the controller.
Controlling air exchange by adjusting the position of the air flap.
Setting or changing access passwords for controller functions.
Function of the class 1 overtemperature safety device.
Overview of safety controller classes and function checks.
Procedure for setting the safety controller limit temperature.
How to interpret and respond to alarm messages.
Overview of available timer functions: Delayed Off, Temp. Dep. Delayed Off, Delayed On.
Setting up and activating the "Delayed Off" timer function.
How to stop or modify the "Delayed Off" timer settings.
Common faults and their solutions for operating personnel.
| Temperature Range | 5 °C above ambient temperature to 100 °C |
|---|---|
| Temperature Display | Digital |
| Power Supply | 230 V, 50/60 Hz |
| Temperature Fluctuation | ±0.3 °C |
| Recovery Time after Door was Opened for 30 seconds | approx. 15 minutes |
| Interior Width | 500 mm |
| Interior Height | 480 mm |
| Number of Shelves (standard/max) | 2/5 |
| Max. Permissible Load | 60 kg |
| Interior Volume | 115 liters |
| Temperature Control | Microprocessor PID control |
| Nominal Power | 600 W |