What to do if my Binder BD 115 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
- SsimpsonjohnAug 31, 2025
The timer may have run off. Try re-programming the timer or turn it off.
What to do if my Binder BD 115 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
The timer may have run off. Try re-programming the timer or turn it off.
What to do if my ED, FD, FED Binder BD 115 chamber has no function and shows alarm message “!TLim”?
This alarm indicates that safety device class 2 has turned off the chamber. Allow the chamber to cool down. Check the settings of the temperature set-point and the safety device class 2, and select a suitable limit value if needed.
What to do if my Binder Laboratory Equipment chamber has no function?
First, check the connection to the power supply. If the overtemperature protective device class 1 has turned off the chamber, disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
What to do if my ED, FD, or FED Binder Laboratory Equipment chamber has no function and shows the alarm message “!TLim”?
The safety device class 2 has likely turned off the chamber. Allow the chamber to cool down. Check the temperature set-point and the settings of the safety device class 2. If appropriate, select a suitable limit value. If the issue persists, further assistance is needed.
Why is my Binder Laboratory Equipment chamber heating permanently and not holding the set-point?
The controller may not be adjusted, so calibrate and adjust it. If the issue persists, the controller might be defective and further assistance is needed.
What to do if Binder BD 115 Laboratory Equipment chamber doesn’t heat up but the controller display is working?
The timer might have run off. Try re-programming the timer or turning it off.
Why is the set-point temperature not reached in my Binder Laboratory Equipment after the specified time?
Several factors could be the reason. Make sure the chamber door is completely closed and that the door gasket isn't defective (replace if needed). The controller may also need calibration and adjustment. Finally, verify the power supply for the correct voltage.
Qualification required for installation, testing, and operation.
Explanation of safety symbols, signal words, and pictograms.
General safety guidelines for installation and operation.
Proper uses, improper uses, and potential misuses of the chamber.
Procedures for unpacking, checking delivery, and safe transport.
Requirements for installation site, storage, and environmental conditions.
Detailed information on connecting the chamber to the power supply.
Procedures for loading the chamber, including weight limits.
Explanation of controller buttons, display, and status symbols.
How the controller shows normal operating status and data.
Procedure to change the controller's display language.
How to set the internal clock and date on the controller.
Method to switch between Celsius and Fahrenheit units.
Entering target temperature and fan speed values via the controller.
Controlling air exchange by adjusting the position of the air flap.
Setting or changing access passwords for controller functions.
Function of the class 1 overtemperature safety device.
Overview of safety controller classes and function checks.
Procedure for setting the safety controller limit temperature.
How to interpret and respond to alarm messages.
Overview of available timer functions: Delayed Off, Temp. Dep. Delayed Off, Delayed On.
Setting up and activating the "Delayed Off" timer function.
How to stop or modify the "Delayed Off" timer settings.
Common faults and their solutions for operating personnel.
| Temperature Range | 5 °C above ambient temperature to 100 °C |
|---|---|
| Temperature Display | Digital |
| Power Supply | 230 V, 50/60 Hz |
| Temperature Fluctuation | ±0.3 °C |
| Recovery Time after Door was Opened for 30 seconds | approx. 15 minutes |
| Interior Width | 500 mm |
| Interior Height | 480 mm |
| Number of Shelves (standard/max) | 2/5 |
| Max. Permissible Load | 60 kg |
| Interior Volume | 115 liters |
| Temperature Control | Microprocessor PID control |
| Nominal Power | 600 W |