EasyManua.ls Logo

Binder BF 260 - User Manual

Default Icon
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Issue 02/2022 Art. No. 7001-0292
Operating Manual
BD (E3.1)
|
Incubators
Avantgarde.Line with natural convection
BF (E3.1)
|
Incubators
Avantgarde.Line with forced convection
ED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with natural convection
FD (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
FED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
and enhanced timer functions
with microprocessor temperature controller
Model Model version Art. No. Model Model version Art. No.
BD 56
BD056-230V 9010/ 9110-0323
ED 260
ED260-230V 9010/ 9110-0339
BD056UL-120V 9010/ 9110-0324 ED260UL-240V 9010/ 9110-0340
BD 115
BD115-230V 9010/ 9110-0325 ED 720 ED720-400V 9010/ 9110-0341
BD115UL-120V 9010/ 9110-0326
FD 56
FD056-230V 9010/ 9110-0303
BD 260
BD260-230V 9010/ 9110-0329 FD056UL-120V 9010/ 9110-0304
BD260UL-120V 9010/ 9110-0330
FD 115
FD115-230V 9010/ 9110-0305
BD 720
BD720-230V 9010/ 9110-0331 FD115UL-120V 9010/ 9110-0306
BD720UL-240V 9010/ 9110-0332
FD 260
FD260-230V 9010/ 9110-0309
BF 56
BF056-230V 9010/ 9110-0313 FD260UL-240V 9010/ 9110-0310
BF056UL-120V 9010/ 9110-0314 FD 720 FD720-400V 9010/ 9110-0311
BF 115
BF115-230V 9010/ 9110-0315
FED 56
FED056-230V 9010/ 9110-0295
BF115UL-120V 9010/ 9110-0316 FED056UL-120V 9010/ 9110-0296
BF 260
BF260-230V 9010/ 9110-0319
FED 115
FED115-230V 9010/ 9110-0293
BF260UL-120V 9010/ 9110-0320 FED115UL-120V 9010/ 9110-0294
BF 720
BF720-230V 9010/ 9110-0321
FED 260
FED260-230V 9010/ 9110-0299
BF720UL-240V 9010/ 9110-0322 FED260UL-240V 9010/ 9110-0300
ED 56
ED056-230V 9010/ 9110-0333
FED 720
FED720-400V 9010/ 9110-0301
ED056UL-120V 9010/ 9110-0334 FED720UL-208V 9010/ 9110-0302
ED 115
ED115-230V 9010/ 9110-0335
ED115UL-120V 9010/ 9110-0336
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: customerservice@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder BF 260

Questions and Answers

  • J
    jeremy38Aug 20, 2025
    What to do if Binder BF 260 chamber has no function?
    • J
      Jennifer NorrisAug 20, 2025
      First, check the connection to the power supply. If that's not the issue, the overtemperature protective device class 1 might have turned off the chamber. Disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
  • A
    Alan StewartAug 22, 2025
    Why is the set-point temperature not reached in my Binder BF 260 after the specified time?
    • T
      Tracie BrownAug 22, 2025
      Several factors could be the reason. Make sure the chamber door is completely closed and that the door gasket isn't defective (replace if needed). The controller may also need calibration and adjustment. Finally, verify the power supply for the correct voltage.
  • A
    Amy ScottAug 25, 2025
    What to do if Binder BF 260 chamber doesn’t heat up but the controller display is working?
    • L
      laurahayesAug 25, 2025
      The timer might have run off. Try re-programming the timer or turning it off.
  • B
    boltongregoryAug 26, 2025
    Why is my Binder BF 260 Laboratory Equipment chamber heating permanently and not holding the set-point?
    • T
      Tina GarciaAug 26, 2025
      The controller may not be adjusted, so calibrate and adjust it. If the issue persists, the controller might be defective and further assistance is needed.
  • B
    brandy23Aug 27, 2025
    What to do if my ED, FD, or FED Binder BF 260 Laboratory Equipment chamber has no function and shows the alarm message “!TLim”?
    • J
      Janet GarciaAug 28, 2025
      The safety device class 2 has likely turned off the chamber. Allow the chamber to cool down. Check the temperature set-point and the settings of the safety device class 2. If appropriate, select a suitable limit value. If the issue persists, further assistance is needed.
  • V
    Victoria LewisAug 29, 2025
    What to do if my Binder BF 260 Laboratory Equipment chamber has no function?
    • M
      Margaret KirbyAug 30, 2025
      First, check the connection to the power supply. If the overtemperature protective device class 1 has turned off the chamber, disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
  • T
    Tiffany MorrisAug 31, 2025
    What to do if my ED, FD, FED Binder Laboratory Equipment chamber has no function and shows alarm message “!TLim”?
    • S
      Seth PattonAug 31, 2025
      This alarm indicates that safety device class 2 has turned off the chamber. Allow the chamber to cool down. Check the settings of the temperature set-point and the safety device class 2, and select a suitable limit value if needed.
  • J
    John BlackSep 2, 2025
    What to do if my Binder BF 260 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
    • M
      Matthew MooreSep 2, 2025
      The timer may have run off. Try re-programming the timer or turn it off.
  • T
    Todd CookSep 4, 2025
    What to do if my BD, BF Binder BF 260 displays alarm message “!TProt”?
    • G
      garycuevasSep 4, 2025
      The alarm message “!TProt” indicates that safety device class 3.1 has responded. Check the settings of the temperature set-point and of the safety device class 3.1.
  • S
    schultzmichelleSep 6, 2025
    What to do if my Binder Laboratory Equipment chamber is heating permanently, and the set-point is not held?
    • D
      datkinsonSep 6, 2025
      The controller may not be adjusted correctly. Calibrate and adjust the controller.

Summary

Safety and Precautions

Personnel Qualification and Responsibilities

Ensures trained staff operate the chamber safely and correctly.

General Safety Guidelines

Covers signal words, symbols, and hazard communication for safe operation.

Intended Use and Risk Management

Defines proper usage, identifies foreseeable misuse, and outlines residual risks.

Installation Procedures

Electrical Connection

Details on connecting the chamber to the power supply, including voltage and fuse info.

Controller Operation and Settings

Temperature and Fan Speed Configuration

Steps for entering set-points for temperature and fan speed.

Overtemperature Protection Systems

Class 1 Overtemperature Device

Explains the Class 1 protective device and its cut-off temperatures.

Safety Controller Configuration

Details Class 2 and 3.1 controllers, set-point adjustment, and alarm management.

Timer Functions Operation

Timer Function Setup and Activation

Covers selecting and configuring Delayed Off, Temp. Dependent Delayed Off, and Delayed On timers.

Data Recording via USB

Network Configuration (Ethernet)

Ethernet Interface Setup

Guides on setting IP address, subnet mask, gateway, and MAC address.

Cleaning and Decontamination

Decontamination and Disinfection

Methods for decontaminating the chamber from hazardous substances.

Maintenance, Troubleshooting, and Repair

Maintenance and Service Intervals

Guidelines for regular maintenance and service requirements.

Troubleshooting Common Issues

Guide to diagnose and resolve typical chamber malfunctions.

Product Disposal

Chamber Decommissioning and Disposal

Procedures for decommissioning and disposing of the chamber in various regions.

Technical Specifications and Data

Binder BF 260 Specifications

General IconGeneral
Temperature RangeAmbient +5°C to 300°C
Voltage230 V
Shelves2
Number of Shelves (Standard/Max)2
DisplayLCD
Rated Voltage230 V
Power Consumption1600 W

Related product manuals