Issue 07/2017 Art. No. 7001-0130
Operating Manual
FP – Drying and heating ovens
with forced convection and program functions
with microprocessor program controller RD3
Model Model version Art. No.
FP 53 (E1.1) FP053-230V 9010-0153, 9110-0153
FP 53-UL (E1.1) FP053UL-120V 9010-0175, 9110-0175
FP 115 (E1.1) FP115-230V 9010-0255, 9110-0255
FP 115-UL (E1.1) FP115UL-120V 9010-0262, 9110-0262
FP 240 (E1.1) FP240-230V 9010-0263, 9110-0263
FP 240-UL (E1.1) FP240UL-120V 9010-0264, 9110-0264
FP 400 (E1.1) FP400-230V 9010-0265, 9110-0265
FP 400-UL (E1.1) FP400UL-120V 9010-0266, 9110-0266
FP 720 (E1.1) FP720-230V 9010-0267, 9110-0267
FP 720-UL (E1.1) FP720UL-120V 9010-0268, 9110-0268
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • J
    Jose DayAug 18, 2025
    What to do if the set-point temperature is not reached in my Binder FP 400?
    • M
      marytateAug 18, 2025
      If the set-point temperature isn't reached after the specified time, there might be a few reasons. First, make sure the chamber door is completely closed. If it is, inspect the door gasket for any defects and replace it if necessary. Also, the controller might need adjustment; try calibrating it.
  • B
    Brian FranklinAug 20, 2025
    What causes my Binder FP 400 chamber to heat permanently and not hold the set-point?
    • C
      Cameron CarrollAug 20, 2025
      If your chamber is heating permanently and not holding the set-point, calibrate and adjust the controller. For other potential causes, contact Binder service.
  • C
    Colleen CastanedaAug 23, 2025
    Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
    • B
      brenda57Aug 23, 2025
      Deviations from the indicated heating-up times can occur if the oven is fully loaded. Try charging the oven with less material or consider allowing longer heating-up times.
  • D
    danthonyAug 25, 2025
    Why is my Binder FP 400 chamber permanently turned off?
    • K
      kimberly17Aug 26, 2025
      If the chamber is permanently turned off, first check the connection to the power supply. Ensure the correct voltage (115V or 230V) is being supplied. Also, check the chamber fuse and replace it if needed. If the fuse responds again, contact Binder service.
  • A
    alisonchapmanAug 28, 2025
    What to do if the red indicator lamp is lit and my Binder Laboratory Equipment chamber is permanently turned off?
    • E
      Ethan MillerAug 28, 2025
      If the red indicator lamp of safety device class 2 is lit and the chamber is permanently turned off, let the oven cool down and then press the RESET button. Check the temperature set-point and the setting of the safety device, and select a suitable limit value if appropriate.
  • T
    Timothy RobertsAug 30, 2025
    What to do if my Binder FP 400 chamber doesn’t heat up, the LED is not lit, but the controller display is working?
    • J
      Jeffrey Kaiser PhDAug 30, 2025
      If the chamber doesn’t heat up, the LED is not lit, and the controller display is working, contact Binder service.
  • M
    Ms. Rebecca RaySep 2, 2025
    What to do if my Binder FP 400 chamber doesn’t heat up, but the LED is lit?
    • B
      bryan97Sep 2, 2025
      If the chamber doesn’t heat up but the LED is lit, contact Binder service.
  • C
    Cindy WalkerSep 4, 2025
    Why is the program duration longer than programmed on my Binder FP 400 Laboratory Equipment?
    • L
      Lisa CoxSep 4, 2025
      If the program duration is longer than programmed, it may be due to inappropriate tolerances. For rapid transition phases, do NOT program tolerance limits to allow maximum heating speed.
  • J
    Jennifer WoodardSep 7, 2025
    How to fix ramp temperature transitions that are only realized as steps in my Binder FP 400?
    • C
      Cheryl WeaverSep 7, 2025
      If ramp temperature transitions are only realized as steps, set the set-point programming type to “Ramp” in the User level.
  • R
    robertsjoyceSep 9, 2025
    Why does my Binder FP 400 Laboratory Equipment program stop one section too early?
    • C
      Chase VegaSep 9, 2025
      If the program stops one section too early, the program line might be incomplete. Define the end value of the desired cycle by adding an additional section with a section time of at least one minute when programming.

Summary

1. Safety

1.4 Type plate

Describes the information provided on the chamber's identification plate.

1.5 General safety instructions on installing and operating the chambers

Provides essential safety guidelines for the installation and operation of the drying and heating ovens.

1.6 Intended use

Defines the approved applications and limitations for the drying and heating ovens.

2. Chamber description

2.1 Chamber overview

Provides a general overview of the drying and heating oven's main components and features.

2.2 Control panel.

Details the layout and function of the standard chamber's control panel.

3. Scope of delivery, transportation, storage, and installation

3.2 Guidelines for safe lifting and transportation

Provides advice and warnings for safely lifting and moving the chamber.

3.4 Location of installation and ambient conditions

Specifies requirements for the installation site and ambient operating conditions.

4. Installation of the equipment

4.1 Electrical connection

Details the procedures and requirements for connecting the chamber to the power supply.

5. Start Up

5.2 Settings at the RD3 program controller

Guides users on initial settings and display functions of the RD3 program controller.

6. Fixed value entry mode

7. Week program editor.

8. Program editor

9. Program start level.

10. User level

11. Example programming for the Week Program Editor

12. Example programming for the Program editor.

13. Behavior at failures

14. Temperature safety devices

14.1 Temperature safety device class 2 (DIN 12880)

Explains the operation, setting, and function check of the Class 2 safety device.

14.2 Temperature safety device class 3.1 (DIN 12880) (option)

Describes the Class 3.1 optional safety device, its function, setting, and checks.

16. Maintenance, cleaning, and service

16.1 Maintenance intervals, service

Specifies recommended maintenance schedules and requirements for authorized service.

16.2 Cleaning and decontamination

Provides detailed instructions for cleaning and decontaminating the chamber.

17. Disposal.

17.3 Disposal of the chamber in the Federal Republic of Germany

Explains specific disposal regulations for Germany under the WEEE directive.

17.4 Disposal of the chamber in the member states of the EC except for the Federal Republic of Germany

Outlines disposal procedures for EC member states outside Germany, following WEEE.

18. Troubleshooting

19. Technical description

19.4 Technical data.

Presents comprehensive technical specifications for various chamber models.

20. EU Declaration of Conformity.

22. Contamination Clearance Certificate

22.1 For chambers located outside USA and Canada

Instructions and form for contamination clearance for non-US/Canada locations.

22.2 For chambers in USA and Canada

Instructions and form for contamination clearance for US and Canada locations.

Binder FP 400 Specifications

General IconGeneral
Internal Volume400 liters
Rated Voltage230 V
Capacity400 liters
Power Supply230 V, 50/60 Hz
Heating Rate5 K/min
MaterialStainless steel
ControllerDigital display with PID microprocessor control