Issue 07/2017 Art. No. 7001-0130
Operating Manual
FP – Drying and heating ovens
with forced convection and program functions
with microprocessor program controller RD3
Model Model version Art. No.
FP 53 (E1.1) FP053-230V 9010-0153, 9110-0153
FP 53-UL (E1.1) FP053UL-120V 9010-0175, 9110-0175
FP 115 (E1.1) FP115-230V 9010-0255, 9110-0255
FP 115-UL (E1.1) FP115UL-120V 9010-0262, 9110-0262
FP 240 (E1.1) FP240-230V 9010-0263, 9110-0263
FP 240-UL (E1.1) FP240UL-120V 9010-0264, 9110-0264
FP 400 (E1.1) FP400-230V 9010-0265, 9110-0265
FP 400-UL (E1.1) FP400UL-120V 9010-0266, 9110-0266
FP 720 (E1.1) FP720-230V 9010-0267, 9110-0267
FP 720-UL (E1.1) FP720UL-120V 9010-0268, 9110-0268
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • M
    Matthew EvansAug 18, 2025
    What to do if the set-point temperature is not reached in my Binder Laboratory Equipment?
    • J
      Jeremy DiazAug 18, 2025
      If the set-point temperature isn't reached after the specified time, there might be a few reasons. First, make sure the chamber door is completely closed. If it is, inspect the door gasket for any defects and replace it if necessary. Also, the controller might need adjustment; try calibrating it.
  • E
    Elizabeth JacksonAug 20, 2025
    What causes my Binder FP 720 Laboratory Equipment chamber to heat permanently and not hold the set-point?
    • B
      Bianca HarperAug 20, 2025
      If your chamber is heating permanently and not holding the set-point, calibrate and adjust the controller. For other potential causes, contact Binder service.
  • Z
    zkleinAug 23, 2025
    Why are there deviations from the indicated heating-up times in my Binder FP 720 Laboratory Equipment?
    • M
      Michele WoodAug 23, 2025
      Deviations from the indicated heating-up times can occur if the oven is fully loaded. Try charging the oven with less material or consider allowing longer heating-up times.
  • A
    Alexa SmithAug 26, 2025
    Why is my Binder FP 720 chamber permanently turned off?
    • J
      justin64Aug 26, 2025
      If the chamber is permanently turned off, first check the connection to the power supply. Ensure the correct voltage (115V or 230V) is being supplied. Also, check the chamber fuse and replace it if needed. If the fuse responds again, contact Binder service.
  • K
    Kenneth BowenAug 28, 2025
    What to do if the red indicator lamp is lit and my Binder FP 720 chamber is permanently turned off?
    • D
      Daniel JacksonAug 28, 2025
      If the red indicator lamp of safety device class 2 is lit and the chamber is permanently turned off, let the oven cool down and then press the RESET button. Check the temperature set-point and the setting of the safety device, and select a suitable limit value if appropriate.
  • G
    Gregory BriggsAug 30, 2025
    What to do if my Binder FP 720 Laboratory Equipment chamber doesn’t heat up, the LED is not lit, but the controller display is working?
    • M
      michaelhowardAug 30, 2025
      If the chamber doesn’t heat up, the LED is not lit, and the controller display is working, contact Binder service.
  • K
    Kara KingSep 2, 2025
    What to do if my Binder FP 720 Laboratory Equipment chamber doesn’t heat up, but the LED is lit?
    • A
      aaron71Sep 2, 2025
      If the chamber doesn’t heat up but the LED is lit, contact Binder service.
  • J
    joel71Sep 4, 2025
    Why is the program duration longer than programmed on my Binder FP 720?
    • K
      katherinescottSep 4, 2025
      If the program duration is longer than programmed, it may be due to inappropriate tolerances. For rapid transition phases, do NOT program tolerance limits to allow maximum heating speed.
  • B
    benjaminlopezSep 7, 2025
    How to fix ramp temperature transitions that are only realized as steps in my Binder FP 720 Laboratory Equipment?
    • C
      Candice ReidSep 7, 2025
      If ramp temperature transitions are only realized as steps, set the set-point programming type to “Ramp” in the User level.
  • R
    rosalesmelindaSep 9, 2025
    Why does my Binder FP 720 program stop one section too early?
    • K
      Kristina PughSep 9, 2025
      If the program stops one section too early, the program line might be incomplete. Define the end value of the desired cycle by adding an additional section with a section time of at least one minute when programming.

Summary

Safety

General Safety Instructions on Installing and Operating the Chambers

Safety guidelines for chamber setup and daily use.

Chamber Description

Control Panel

Details of the interface for operating the chamber.

Scope of Delivery, Transportation, Storage, and Installation

Location of Installation and Ambient Conditions

Requirements for chamber placement and operational environment.

Installation of the Equipment

Electrical Connection

How to connect the chamber to the power supply.

Start Up

Settings at the RD3 Program Controller

Initial configuration of the RD3 program controller.

Fixed Value Entry Mode

Week Program Editor

Program Editor

Selecting Between Set-Point Ramp and Set-Point Step

Choosing temperature transition methods: ramp or step.

Programming with Setting Ramp (Default Setting)

How to program gradual temperature changes.

Programming with Setting Step

How to program sudden temperature changes.

Set-Point Entry for Program Operation

Detailed steps for entering program values.

Deleting a Program Section

Procedure for removing program segments.

Example Programming for the Week Program Editor

Example Programming for the Program Editor

Desired Time Function

Defines target temperatures for different periods.

Proceeding in Detail

Step-by-step guide for program editor setup.

Behavior at Failures

Behavior After Power Failure

How the chamber behaves after power interruptions.

Alarm Messages

Explanation of error messages and buzzer functions.

Temperature Safety Devices

Temperature Safety Device Class 2 (DIN 12880)

Description and function of the class 2 safety device.

Temperature Safety Device Class 3.1 (DIN 12880) (Option)

Description of the optional class 3.1 safety device.

Maintenance, Cleaning, and Service

Cleaning and Decontamination

Procedures for cleaning and disinfecting the chamber.

Troubleshooting

Technical Description

Technical Data

Detailed specifications for chamber models.

Binder FP 720 Specifications

General IconGeneral
BrandBinder
ModelFP 720
CategoryLaboratory Equipment
LanguageEnglish