2) Hard connection belongs to A2 safety cabinet’s special operation method. If you need to use this
method, please contact us.
3)Flow alarm system instruction
Flow alarm system is an independent software package programmed in advance to monitor the
abnormal of air volume. It alarms when the air volume exceed normal value.
2.1.11Technical Parameter
Model
Technical parameter
BSC-4FA2-GL
BSC-6FA2- GL
Rated voltage AC
220V±10% 110V±10%
Rated frequency
50 Hz 60Hz
External Size(W*D*H)
1383×775×2295mm
1873×775×2295 mm
Working Zone
Size(W*D*H)
1210×600×660mm
1700×600×660 mm
Rated current
9A
9A
Total Airflow Volume
590m3/h
825m3/h
UV Lamp Consumption
30W
40W
Fluorescent lamp
Consumption
18W*2
22W*2
Down flow Velocity
0.33m/s
Inflow Velocity
0.53m/s
HEPA Filter
99.999(Diameter:0.3μm)
Noise
NSF 49≤61dB / EN 12469≤58dB
Notes: (1) Electric consumption power including power which operation area needs to load(Loading
no more than 500W)
(2) Our company reserves the right of product design changes, if there are any design change,
without prior notice.