6.2 INJECTION
WARNING
If you are not certain that you injected your insulin, do not start or repeat your injection.
Monitor your blood glucose as instructed by your healthcare provider.
6.2.1 Check the insulin dose to inject according to the prescription and the type of
medication.
6.2.2 Turn the button to select the number of units to inject. The dose indicator should
line up with the dose to inject.
6.2.3 Press on the injection button until you return to ZERO (complete dose injection)
6.2.4 Keep the injection button pressed in
and the needle in the skin:
- While the Mallya’s green LED is flashing,
-
Until Mallya emits an audible signal
(BEEP) signaling the end of the injection.
6.2.5 You can release the pressure on the button and then remove the needle from the
skin when the green LED stops flashing and the “BEEP” signaling the end of
injection is emitted.
WARNING
If you do not see ZERO (“0”) in the pen dose window: it indicates that the full dose has not
been injected.
If the above instructions are followed, the Mallya application will automatically
detect an injection:
The information corresponding to the injected dose is displayed:
Injected dose, day and time.