This test protocol can be used to compare patients of similar age ranges in a normative database.
Test Duration: 20 seconds
Level: 12 to 8
Stance Type: Bilateral
A minimum of three test trials should be used to avoid excessive balance deviations.
The patient’s performance is noted as a stability index. Test results are compared to age depend-
ent normative data. Scores higher than normative values suggest further assessment for lower
extremity strength, proprioception and vestibular or visual deficiencies. Poor balance is a major
contributor to falls. Strength, particularly high speed, and proprioceptive training have demon-
strated positive results in balance improvement.
1
For information on normative data referenced in the Predictive Values Report, see Appendix C.
For information comparing the original Balance System Protocol for Fall Risk Balance Testing to
the Balance System SD Protocol for Fall Risk Balance Testing, see note and chart page B-1.