Fetal Monitor FC-700 Operating Manual
< Final Report >
- Period : Time during which CTG is analyzed
- Average Baseline FHR(bpm) : Mean number of Baseline FHR during above period
- Number of UC(Frequency / h) : Number of UC and frequency of UC per hour during the
period
- Number of ACC(Frequency/h): : Number of Acceleration and frequency of Acceleration per
hour during the period
- Number of DEC(Frequency /h) : Number of Deceleration and frequency of Deceleration
per hour during the period
- Late DEC : Number of Late Deceleration during Deceleration
- Early DEC : Number of Early Deceleration during Deceleration
- Variable DEC : Number of Variable DEC during Deceleration
- TACHY : MODERATE(>160bpm) : Time in minute during which FHR is 160~190 bpm
- TACHY : SEVERE(>190bpm) : Time in minute during which FHR is greater than 190
bpm
- BRADY : MODERATE(<110bpm) : Time in minute during which FHR is 110~90 bpm
- BRADY: SEVERE (<90bpm) : Time in minute during which FHR is smaller than 90bpm
CTG REPORT (PERIOD : 0 - 15 MIN 0 SEC)
AVERAGE BASELINE FHR(BPM) : 123 bpm
NUMBER OF UC(FREQUENCY) : 5/ 20.0/h
NUMBER OF ACC(FREQUENCY) : 0/ 0.0/h
NUMBER OF DEC(FREQUENCY) : 5/ 20.0/h
LATE DEC : 0
EARLY DEC : 5
VARIABLE DEC : 0
TACHYCARDIA : MODERATE(>160BPM) : 0.0 min
SEVERE(>190BPM) : 0.0 min
BRADYCARDIA : MODERATE(<110BPM) : 0.0 min
SEVERE(<90BPM) : 0.0 min