Fetal Monitor FC-700 Operating Manual
2) Recording Setup
It is a mode to control the setup values of printing speed, grid indication, contrast, automatic
NST function, and automatic fetal movement detection.
How to Operate the Recording Setup Mode
Press the Reference key for over two seconds to change into the recording setup mode.
Press the Reference key for a short time at the recording setup mode to move to the next
item.
Press the VOLUME / key to change the set-up value.
Press the Record key to store the set-up value and then return it to the basic state.
Press the Alarm key to cancel the set-up value and then return it the basic state.
How to Set the printing Speed, Grid Indication, and Contrast
To set the printing speed, press the Reference key for over two seconds and then for a
short time to indicate the following. It indicates that the (output is at a speed of 3
centimeters a minute now.) printing speed is 3 cm/min now.
The output speed can be set at one of 1, 2, and 3 cm/min, and may be changed with the
Volume Up/Down Key.
Thermal paper for a facsimile can be used for FC-700 besides those supplied. To use those
for a facsimile, press the Reference key for over two seconds and then for a short time to
indicate the following.
Select 0 for sheets supplied by this company and 1 for those for a facsimile, and use the
Volume Up/Down Key to change it. If the set-up value is 1, Grid is printed at a sheet for a
facsimile with a signal.
Note
When use a paper just for fetal monitor, set-up value must be “0”..