Disposal of your old appliance
1. When this crossed out wheeled bin symbol is attached to a product it
means the product is covered by the European Directive 2002/96/EC.
2. All electrical and electronic products should be disposed of separately
from the municipal waste stream via designated collection facilities
appointed by the government or the local authorities.
3. The correct disposal of your old appliance will help prevent potential negative
consequences for the environment and human health.
4. For more detailed information about disposal of your old appliance, please
contact your city office, waste disposal service or the shop where you
purchased the product.
5. instructions indicated the responsibility of the RESPONSIBLE ORGANIZATION to comply
with international, regional or national regulations concerning environmental protection
6. Maintenance of equipment
1) When DOP Probe does NOT work, it is connected to the other channel or exchange the DOP
Probe. Also equal UC Probe.
2) If equipment does NOT work, you have to make a phone call service person of Bionet.
For more information, see the Service Manual.
2) Regular Inspection
Perform the periodical safety inspection on FC-700 once a year. For the inspection details, see
the service manual provided by Bionet
3) Error Message
A. If the Doppler probe comes off from the connector during monitoring, information sound
(“Ding-Dong”) rings and the error message(Er1) is displayed. To solve this problem,
connect the Doppler probe or press the VOLUME UP/DOWN key simultaneously.
B. If the paper is used up during printing, information sound (“Ding-Dong”) rings and the error
message (Er2) is displayed. To solve this problem, insert the paper or press the VOLUME
UP/DOWN key simultaneously.