In case Gel contacts internal electronic device, Defective probe may
cause electrical shock.
Prior to each use, visually inspect the probe lens and case area for
cracks, cuts, tears, and other signs of physical damage.
DO NOT use a probe which appears to be damaged until you verify
functional and safe performance.
Perform a more thorough inspection, including the cable, and
connector, each time clean the probe.
DO NOT kink, tightly coil, or apply excessive force on the probe cable.
Insulation failure may result.
Warning statement for class i me equipment indicating: “warning: To
avoid risk of electric shock, this equipment must only be connected to
a supply mains with protective earth”
“Do not modify this equipment without authorization of the
manufacturer”
“warning: If this equipment is modified, appropriate inspection and
testing must be conducted to ensure continued safe use of equipment”
Do not to touch signal input, signal output or other connectors, and the
patient simultaneously.
Refer servicing to qualified personnel of Bionet Co., Ltd.
Ultrasound can produce harmful effects in tissue and potentially result
in patient injury. Always minimize exposure time. and keep ultrasound
levels low when there is no medical benefit.
From probe bottom to 2~3 cm, waterproof IPX is possible. Do NOT
immerse the probe bottom into any liquid beyond 2~3 cm from probe
bottom. Never immerse the probe connector into any liquid.