Rev.2.41 49
Fetal Monitor FC-700 Operating Manual
Acoustic Output Tables
MC65R1S Pulsed Doppler Mode
Index
MI
TIS
TIB
TIC
scan
Non-scan
Non-scan
Aaprt<= 1
Aaprt > 1
Global Maximum : Index Value
0.014874
-
0.010567
-
0.166265
0.178513
IEC
FDA
Unit
Associated
Acoustic
Parameter
pra
pr.3
(MPa)
0.0148732
P
Wo
(mW)
-
43.7
43.7
43.7
min of [Pα(zs),
Ita,α(zs)]
min of [(W.3(Z1),
ITA.3(z1)]
(mW)
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
12.226
12.226
z at max. Ipi,α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
3.28258
ƒawf
fc
(MHz)
0.999891
-
0.999891
-
0.999891
0.999891
Dim of Aaprt
X
(cm)
-
6.12
-
6.12
6.12
Y
(cm)
-
6.12
-
6.12
6.12
Other
Information
td
PD
(μsec)
93.3692
prr
PRF
(Hz)
4000
pr at max. Ipi
pr@PIImax
(MPa)
0.0202734
deq at max. Ipi
deq@PIImax
(cm)
3.2769
Focal Length
FLX
(cm)
-
0
-
0
FLY
(cm)
-
0
-
0
Ipa,α at max. MI
IPA.3@MImax
(W/cm^2)
0.00521243
Operating
Control
Conditions
Frequency
(MHz)
1.0
-
1.0
-
1.0
1.0