Fetal Monitor FC-700 Operating Manual
Power
frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Voltage dips,
short
Interruptions
and
Voltage
variations
on power
supply
input lines
IEC 61000-4-11
<5% Uт
(>95% dip in Uт)
for 0.5cycle
40% Uт
(60% dip in Uт )
for 5 cycle
70% Uт
(30% dip in Uт)
for 25 cycle
<5% Uт
(<95% dip in Uт )
for 5 s
<5% Uт
(>95% dip in Uт)
for 0.5cycle
40% Uт
(60% dip in Uт )
for 5 cycle
70% Uт
(30% dip in Uт)
for 25 cycle
<5% Uт
(<95% dip in Uт )
for 5 s
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the FC-700 system requires
continued operation
during power mains
interruptions, it is
recommended that the
FC-700 system be
powered from an
uninterruptible power
supply or a battery
Note: Uт is the a.c. mains voltage prior to application of the test level.
The FC-700 system is intended for use in the electromagnetic environment specified
below.
The customer or the user of the FC-700 system should assure that it is used in such an
environment
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Portable and mobile RF
communications equipment
should be used no closer to any
part of the FC-700 system,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance