Performing a Quality Control Test
Performing a Quality Control Test
9. Tightly replace the cap on the Control Solution bottle.
10. The Control Solution result appears. Compare your Quality Control Test result to the Control Solution
Range printed on the Blood Glucose Test Strip GS550 vial label.
Rightest
Rightest
®
®
1. Take one Blood Glucose Test Strip GS550 from vial and re-cap the vial cap
immediately.
2. Insert the Blood Glucose Test Strip GS550 with view window, facing up, into test
strip port.
3. While the blood drop symbol flashing, press and hold the main button for over 3 seconds until
the " " symbol appears.
4. You will see blinking " " symbol and " " symbol symbol on the screen, prompting you to
apply Control Solution.
5. Shake the bottle of Control Solution well before opening the cap. Then open the
bottle and put the cap on the table.
6. Drip a drop of Control Solution on the top of the cap.
7. Gently touch sample entry of the strip with the Control Solution on the top of the
cap.
Rightest
Rightest
Rightest
Rightest
Rightest
Rightest
8. When you hear a beep (if volume is turned on ) please wait for the test result. The screen will
display the countdown number from 5.
®
®
®
®
®
®
OPEN
1
2
4
5
6
CLOSE
9
PRECAUTION
- Your Control Solution results will not be calculated for average reading but still can be recalled. The Control
Solution Test result will be shown with " " symbol on the screen.
- Our suggestion environment for Control Solution Test is 59-104(15-40).
- Before " " and " " appears, please don't touch the Control Solution to the sample entry on strip
because the Blood Glucose Meter GM550 is still in an internal check. If you do so, the meter will show
" " and " ". Accompanied by beeps (if volume is turned on).
- Don't drip the Control Solution to sample entry of the strip directly. The reagent on strip
might be sucked into the bottle of Control Solution and might cause the
degeneration of Control Solution. Doing this might contaminate the meter via the test
strip port as well.
- Don't touch the tip of the Control Solution bottle. If you have touched it, please clean up with water.