Limitations
Specification
- Store the Rightest Blood Glucose Test Strips GS550 in the original capped vial at temperatures between 39 to 86( 4 to
30 ) and relative humidity below 90%. Do not freeze.
- Rightest Blood Glucose Test Strips GS550 are designed for using with capillary whole blood samples. Do not use serum or plasma
samples.
- Inaccurate test results may be obtained at high altitude more than about 10000 feet ( 3048 meters ) above sea level.
- Hematocrit below 30% may cause higher results, and hematocrit above 60% may cause lower results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Rightest Blood Glucose Monitoring System GM550 has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Uric acid 10.0 mg/dL, L-Dopa 3 mg/dL, Dopamine 2 mg/dL, Ascorbic acid (Vitamin C) 5 mg/dL
®
®
®
NOTE
- Acetaminophen, Ibuprofen, Methyldopa, Salicylic Acid, Tetracycline, Tolbutamide, Bilirubin-unconjugated, Cholesterol,
Creatinine, Triglyceride, Maltose, Xylose, Galactose and Lactose ( when occurring in normal blood or normal therapeutic
concentrations ) do not significantly affect results. However, abnormally high concentrations in blood may cause inaccurate
results.
- Sodium Fluoride/Potassium Oxalate at low glucose levels may result in inaccurate blood glucose reading. Sodium
Fluoride/Potassium Oxalate should not be used as an anticoagulant with this device.
Measurement Technology
Measurement Calibration
Sample
Minimum Sample Volume
Measuring Range
Test Time
Memory Capacity
Power Saving
Operating Temperature
Operating Relative Humidity
Oxidase Electrochemical Sensor
Plasma
Capillary whole blood
1.0 microliters
20-600 mg/dL
5 seconds
500 blood glucose test results with date and time
Turn off automatically 2 minutes after last user action
Press the " " button for 4 seconds.
50 ~ 104 (10 ~ 40)
10 - 90%