Error messages and Trouble shooting
Error messages and Trouble shooting
Battery Error
1. The " " symbol is blinking when the battery power is low. Please change the meter battery as soon as you can. You
can still do the test.
2. The " " and " " symbols are blinking when the battery is too low. Meter can not do the strip test. Please change
the battery immediately. meter
1a
2
NOTE
- When Blood Glucose Meter GM550 backlight has no function, you cannot see the backilght, please
change your backlight battery.
- When the backlight battery is running out of power, your Blood Glucose Meter GM550 still works
properly but without backlight function.
Rightest
Rightest
®
®
Er1 - The inserted test strip has been used or damaged. Please use a new test strip
from vial.
Er2 - Meter has malfunctioned. Do the Quality Control Test or reinstall the batteries to
check if the meter works properly.
Er3 - Signal transmission is disrupted, repeat the test.
Er4 - Applied blood volume is insufficient, please repeat the test with a new test strip.
Above, if error screen still appears contact your local Bionime Customer Service.