Precaution
Precaution
- Before using the Rightest Blood Glucose Monitoring System GM550 to test your blood glucose, please read all the
instructions and practice the tests including quality control test ( Refer to page 36.)
- Please do the quality control test regularly to make sure the test results are accurate .
- The Rightest Blood Glucose Meter GM550 can only be used with the Rightest Blood Glucose Test Strips GS550.
The use of any other brand Strips should not be used under any circumstances. The use of other brands of strips may
give inaccurate results.
- The Rightest Blood Glucose Monitoring System GM550 is intended for in vitro diagnostic use only. The testing result is
calibrated to be plasma equivalent for test with fresh capillary whole blood samples from the fingertip, palm and forearm.
- The Rightest Blood Glucose Monitoring System GM550 is intended for self-testing. It should not be used to diagnose
diabetes mellitus.
- Rightest Blood Glucose Monitoring System GM550 has not been validated for use on neonates. Therefore, it's not
intended for use of neonates.
- Rightest Blood Glucose Monitoring System GM550 hasn't been used for arterial blood testing yet, so it is still not
suitable to be used for arterial blood testing so far.
- Do the test at least 30 minutes after while entering another site with different ambient temperature.
- Pay attention to the environment protection when disposal of batteries.
- Using appliance in tropical and / or temperate climate.
®
® ®
®
®
®
®
Please take a minimum of 1.0µL to do the test on Rightest Blood Glucose Monitoring System GM550. Blood sample
size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 1.0µL might cause inaccurate test result or might not start on meter measurement. In this case,
repeat the test with a new test strip.
®
PRECAUTOIN
The Rightest Blood Glucose Meter GM550 will be set to mg/dL by default when sold in the United States.
When your display shows " mmol/L " during setting or test, please contact customer service. Use of the wrong
unit of measure may cause you to incorrect treatment.
®
Sample Size Example
- Not to be exposed dripping or splashing.
- The minimum blood sample size of Rightest Blood Glucose Monitoring System GM550 is 1.0µL : ( )
®
1.0µL 1.5µL 2.0µL 3.0µL 4.0µL