411829-L, Operation April 2012 Page 4-18
4
To reuse the peristaltic pump tube used in the previous run (not recommended
see the note
above), flush and fill the tubing with a solvent suitable for the next purification:
1. Place the pump’s inlet line into the solvent of your choice.
2. Place the pump’s outlet line into a waste reservoir.
Figure 4-13. Sample Loading Pump
3. Select the Solvents tab in the right-hand panel (in the software) and then the Sample
Loading Pump tab.
4. To start the pump, press Start. Adjust the flow rate using the Speed slider.
5. When you are done, press Stop.
Figure 4-14. Sample Loading Pump Tab
6. Connect the pump’s outlet line to the cartridge’s inlet.
7. Place the pump’s inlet line into the sample.
Outlet line Inlet line