38 Technical Data
Power Supply
Type Primary clocked broadband power supply
Mains voltage [V],
Frequency [Hz]
100 115 V ± 10% / 60 Hz / 1.2 A / AC
220 230 V ± 10% / 50 Hz / 0.6 A / AC
Safety class I, DIN EN 60601-1, Sec. 5.1
Fuse [A] 3.15 surge-proof
Power [ W ] Continuous power: 100
Short-term maximum: 140
Charging Circuit
Type Switch mode charging circuit corresponding to battery
type
Safety switch To prevent overloading and excessive temperature
Recharging time
[hrs]
4; charging with 1/3 C
Internal Printer
Power [ W ] Controller and on standby: 1
Thermoline and motor: 50
Type of printer Thermal printer
Resolution
[dots/mm]
8
Paper width [mm] 112
Printing width
[mm]
104
Feed rate [mm/s] 5, 10, 25, 50, for graphics as well
Feed rate
tolerance [%]
±2,5;
Maximum absolute error over 100 mm printout