GROUND TERMINAL
1-2 Technical Specification
Important Note: All accuracies apply to 5% to 95% full scale of each range
- DC VOLT
Range Resolution Accuracy Impedance
2 V
20 V
200 V
100 uV
1 mV
10 mV
±(0.05% + 5d)
1000V
100m V ±(0.1% + 5d)
10 M
1200V
1200V
100m V ±(0.3% + 5d)
- AC VOLT
Range Resolution Frequency Accuracy Impedance
2V
100 uV
40Hz to 60Hz
±(0.75% + 10d)
60Hz to 1kHz
±(1.5% + 10d)
1kHz to 5kHz
±(2.5% + 10d)
10 M
5kHz to 10kHz
±(5.0% + 10d)
20V
1 mV
40Hz to 60Hz
±(0.75% + 10d)
60Hz to 1kHz
±(2.5% + 10d)
1kHz to 5kHz
±(5.0% + 10d)
5kHz to 10kHz
±(10% + 10d)
200V 10 mV 40Hz to 50Hz
±(1.0% + 10d)
50Hz to 60Hz
±(0.75% + 10d)
60Hz to 1kHz
±(2.5% + 10d)
1kHz to 5kHz
±(5% + 10d)
1000V 100 mV 40Hz to 50Hz
±(1.0% + 10d)
50Hz to 60Hz
±(0.75% + 10d)
60Hz to 400Hz
±(2.5% + 10d)
400Hz to 1kHz
±(5% + 10d)