V3A | 08/2018
41
© by Blum-Novotest
Measuring sequence
NOTICE
Risk of collision
If a M-function is already active (e.g. M46 probe system On), no further flash is allowed
to be given when the same M-function is repeated as long as the
ERROR signal gives
the message "probe system ready".
Otherwise the probing system is not switched on with M46 but switched off!
If the probe is not deflected for more than 5 minutes, it is additionally set to
Standby.
Realise activation / deactivation of probe with two different M-Codes.
Example:
M46: probe system on