EasyManua.ls Logo

BLX v100 positioner - Calibration

BLX v100 positioner
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.7 Calibration
The V100 is delivered factory
calibrated 0-100 % ±1%.
Calibration procedure
Zero position
1. Set 0% input signal.
2. Wait until the valve has adjusted.
3. Adjust the zero position by turning
the zeroing screw(3), with a screw-
driver from the outside or by using
a slot(3a) on the yellow wheel.
Range
4. Increase to 100% input signal.
5. Wait until the valve has adjusted.
6. Adjust the range by turning the red
range nut(6).
Check the zero position
Make fine adjustments if necessary.*
* With Split Range, where zeroing can be
done by a signal other than 0%, the
steps
1-6 must be repeated until the desired
setting has been reached.
11
V100 Positioner