Service Manual for RESmart
®
BPAP V1.1
3/18
a) Select the upgrade file which has same suffix letter as the machine version. (For example
Version 1.88-TN should choose file name 1.98-TN)
b)
Select the right serial port, default is
COM1.
3.
Click “OK”, display as be
low:
Power on the BP
AP device. If the device is powered on from beginning, please power off the device
and power on again after 5 seconds or more.
4.
After the device is turned on, the following will di
splay:
Click “Upgra
de” to start. If success, the right version number will be displayed on BPAP device
screen when every power on.
Note: If failed during upgrading, please repeat from above 1-4 steps.
If you select an incorrect upgrade file (for example: apply a CPAP upgrade file to a BPAP device), the
device will display “Error 08” and cannot work. In this case, apply correct software and upgrade
again.
3.2 Hidden Function
1. Humidifier matching
Assemble humidifier on main device, without water chamber
. Enter the maintenance menu, and set
Ramp=40, then press Heated Humidifier Button when the Init P appears. Thus the humidifier is
matched. When ‘OK’ appears on screen, humidifier matching succeeded.
2.
Reset (A
vailable on version 1.58 and
later)
Press and h
old the Ramp Button until the user menu appears (about 3 seconds) when the device is
on standby, press +/- User Buttons to access the Date Setting, set ‘YYYY/MM/DD’=’2097/07/01’,
then press ‘Pressure Start/Stop Button’ when ‘Hour’ setting appears, thus all device settings and
therapy records (doesn’t include Use Days and Use Time) will be reset.
If set ‘Minute’=’16’ during the above procedure, then not only device settings and therapy records,
but also patient information will be reset.
If set ‘Minute’=’26’ during the above procedure, then not only device settings, therapy records and