Service Manual for RESmart
®
BPAP V1.1
4/18
patient information, but also Use Days and Use Time will be reset.
3. Pressure c
alibration (Available on version 1.58 and later)
Connect pressure meter to the air outlet of the device, in the Date and Time Setting menu, set
‘YYYY/MM/DD’=’2098/08/18’, press ‘Pressure Start/Stop Button’ when Hour Setting appears to start
the Pressure Calibration. The device output pressure is based on 20 hPa (cmH
2
O), and a number will
appear on the screen, when the output pressure is stable, if the value on the pressure meter is not 20
hPa (cmH
2
O), press ‘+/-’ User Buttons to adjust the output pressure , finally press ‘Pressure
Start/Stop Button’ to finish the Pressure Calibration.
4. Malfunction and Countermeasure
4.1 Error Code Table and Information
Error Index Problem Causing
Error1 Over temperature inside or
Temperature sensor fault
Malfunction on main board
Error2 Motor stop Malfunction on motor or main board
Error3 Motor low speed Malfunction on motor or main board
Error4 Low temperature inside Malfunction on main board
Error7 Pressure sensor fault Malfunction on main board
Error8 Wrong software version Software
Error9 Host parameter error Malfunction on main board
Error12 Over temperature inside or
flowrate sensor fault
Malfunction on main board
Error13
Temperature below 0℃ or
flowrate sensor fault
Malfunction on main board
Error14 The motor is over
temperature or motor
temperature sensor fault
Malfunction on motor or
Motor temperature sensor
Error15 Motor temperature sensor
fault
Malfunction on motor temperature
sensor