OZONE ANALYZER BMT 964 Manual, Rev. 04/2021
30
In case of excessive dirt (e.g. Dirty Warning or Dirty Alarm as described on page 26), the cuvette
should be cleaned. We recommend to send in the analyzer to the manufacturer or representatives
(see chapter Troubleshooting).
Warning: The cuvette may not be flushed with fluids, namely water, for cleaning, as this could
damage the pressure sensor.
The UV lamp is the only part of the instrument which undergoes wear. Life expectancy of the
lamp is several years. When Lamp Low Warning is activated (see page 25) replacement of the
lamp should be planned for the next 4 to 8 weeks. Replacement by the user is possible but we do
not recommend it. The instrument should rather be sent back to us for replacement of the
UV lamp and for recalibration with the new lamp.