BLT310 Product Manual
5
3.Water Cooling Interface
1A
2A
Water cooling water
outlet
Water cooling water
Caution: Installation connection: water-cooled inlet port 2A, water-cooled outlet port 1A
Pure water or drinking water shall be used for coolant.
In order to prevent the water in the cooler from growing molds which can cause blockage of the pipeline, it is
recommended to add ethanol(CH3CH2OH) when filling pure water, and the amount of ethanol
(CH3CH2OH) added is 10% of the pure water.
When the ambient temperature around the equipment is -10C~0C, glycol solution with a volume ratio of
30% must be used and replaced every two months.
When the ambient temperature around the equipment is lower than -10C, a water cooler with dual system
function must be used and the cooling system must be guaranteed to run uninterruptedly.
Recommended cooling water setting values: cooling water pressure ≤ 5bar (0.5Mpa), water flow rate ≥
2.0l/min.
Please refer to the dew point table for the cooling water temperature setting to prevent condensation on the
optical components.