Boonton 4540 Series RF Power Meter
Remote Operation
DISPlay:PULSe:TSPAN
(Note: This command is supplied for compatibility with the Model 4530 Power Meter)
Description: Set or return the horizontal time span of the display in pulse mode. Time span = 10*
Time/Division. Optional units: minutes.
Syntax: DISPlay:PULSe:TSPAN <numeric_value>
Argument: <numeric_value> = 10e-8 to 100 sec in a 1-2-5 sequence,
300 s,
10, 20, 50, 100, 300, 600 min
Valid Modes: Pulse mode
Note: There are separate time bases for the modulated mode and the pulse mode. The
arguments selected are saved and restored independently by mode.
DISPlay:SCREensaver:BRIGhtness
Description: Set or return the display backlight brightness in percent when dimmed..
Syntax: DISPlay:SCREensaver:BRIGhtness <numeric_value>
Argument: <numeric_value> = 0 to 100
Valid Modes: All
DISPlay:SCREensaver:STATe
Description: Set or return the state of the screensaver function. When ON, the screensaver dims the LCD
backlight to a specified intensity after a specified period of keyboard inactivity. GPIB activity
does not disturb display backlighting unless DISPlay:SCREensaver commands are sent.
Syntax: DISPlay:SCREensaver:STATe <character data>
Argument: <Boolean> = { 0, 1, OFF, ON }
Valid Modes: All
DISPlay:SCREensaver:TIMe
Description: Set or return the time in minutes of no activity that dims the display backlight.
Syntax: DISPlay:SCREensaver:TIMe <numeric_value>
Argument: <numeric_value> = 1 to 180
Valid Modes: All