EasyManua.ls Logo

Boonton 4541 - Fetch:interval:minimum; Fetch:interval:pkavg; Fetch:key; Fetch:marker:average

Default Icon
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Boonton 4540 Series RF Power Meter
Remote Operation
FETCh:INTERval:MINimum?
Description: For the specified channel, return the minimum instantaneous power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax: FETCh[1|2]:INTERval:MINimum?
Returns: CC, minimum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
FETCh:INTERval:PKAVG?
Description: For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.
Syntax: FETCh[1|2]:INTERval:PKAVG?
Returns: CC, peak-to-average ratio of power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
FETCh:KEY?
Description: Return the key code of the last key depressed; e.g. ESC = 8.
Syntax: FETCh:KEY?
Returns: CC, key code
Where CC is the measurement condition code.
Valid Modes: All
FETCh:MARKer:AVERage?
Description: For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.
Syntax: FETCh[1|2]:MARKer[1|2]:AVERage?
Returns: CC, average power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.