EasyManuals Logo

Boston Scientific Vercise DB-5572-1A User Manual

Boston Scientific Vercise DB-5572-1A
1162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
MRI Mode
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 41 of 1157
en
Impedances Out Of Range Screen
The impedances must be within the acceptable range before MRI Mode is enabled. If the impedances are
not within the acceptable range, the Remote Control will display an error message.
1. Press to continue.
2. The Remote Control displays a new message instructing the user to review the MRI scan risks related
to abnormal impedances. Review the “Impedance Out of Range” section in the ImageReady
TM
MRI
Guidelines before proceeding. Press
to continue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DB-5572-1A and is the answer not in the manual?

Boston Scientific Vercise DB-5572-1A Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DB-5572-1A
CategoryMedical Equipment
LanguageEnglish

Related product manuals