Stimulator Battery Messages
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 43 of 1157
en
Stimulator Battery Messages
ERI (Elective Replacement Indicator) Screen (For Non‑Rechargeable Stimulators
Only)
This screen displays when the implanted non-rechargeable Stimulator is nearing end of service.
Stimulation is provided during the ERI period. Changes made to the stimulation will not be saved, and
stimulation will not be available soon. Batteries that have lasted one year or more without entering the ERI
period will have a minimum of four weeks between entering the ERI period and reaching end of battery
life. Patients should be advised to contact their healthcare provider to report this message screen. Press