Vercise DBS Remote Control 4 Directions for Use
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 6 of 1157
en
Establishing Communication Between Stimulator and Clinician Programmer
The Remote Control can be used to establish communication between the Stimulator and Clinician
Programmer (CP) for device programming. The Remote Control must be linked to the Stimulator to
perform this function.
From the Lock screen, press and hold the
button (approximately 5 seconds) until the screen displays
“Stimulator Ready for CP Connection.”
This will place the Stimulator into CP Mode (Pairing Mode) so that a CP may connect to it. Then proceed
to connect the CP to the Stimulator. See the “CP Mode: Stimulator” section of this manual for additional
information.