4. Preparation
4.1. Setting up the argon device
NOTE
Electromagnetic fields are generated during normal use of the argon and HF
devices. This can adversely affect other devices.
Ensure that no electronic devices are placed in the vicinity of the argon and
HF devices.
Argon and HF devices may only be used in rooms used for medical purposes that
fulfil the requirements of DIN VDE 0100-710.
If the argon device has previously been stored or transported at temperatures
below +10 °C, it will take approximately three hours to adjust to room temperature.
Before using the device for the first time, ensure that the combination of the argon
device and the electrosurgery device has been functionally tested and the
person(s) responsible for the operation of the devices has/have been instructed by
the manufacturer or the supplier.
The HF device, ARC PLUS device and argon bottle are usually fitted on an
ARC CART equipment trolley with a gas bottle holder.
Figure 4-1: ARC CART