997783 Operating manual
47
8 Troubleshooting
Problem Possible cause Corrective action
Pipette broken off in adapter Pipette was damaged or
pushed into the adapter
with too much force
Remove glass fragments
with tweezers; replace
adapter if necessary, see .
Instrument suction system not work-
ing/pipette dripping
Suction system damaged Replace suction system,
see .
Liquid in suction system Adjustment wheel turned
too quickly
Clean or disinfect suction
system and replace if neces-
sary, see .
Ejector button does not work Ejector button not posi-
tioned properly
Remove clip. Hold the in-
strument loosely and briefly
press the ejector button.
Slide clip on, see .
Pipette is ejected from the adapter Adapter is too moist or
greasy
Clean or disinfect suction
system and allow to dry; re-
place if necessary, see .
NOTICE
A visual inspection of the device is often sufficient to detect a defective
part. This can then be easily replaced with a spare part. The advantages
of self-repair are obvious: Cost and time savings.