micro-Pipettierhelfer
micro Pipette Controller
Gebrauchsanleitung
Bitte vor Gebrauch sorgfältig lesen und alle
Bedienungs- und Sicherheitshinweise befolgen!
BRAND GMBH + CO KG
Otto-Schott-Str. 25
97877 Wertheim
Germany
Tel.: +49 9342 808-0
Fax: +49 9342 808-98000
E-Mail: info@brand.de
Internet: www.brand.de
Operating Manual
Please read carefully before use, and follow
all operations and safety instructions!

Other manuals for BRAND micro

Questions and Answers

  • T
    tinahardinAug 16, 2025
    What to do if there is liquid in suction system of BRAND micro?
    • R
      Rachel SpencerAug 16, 2025
      If you find liquid in the suction system of your BRAND Laboratory Equipment, it's likely because the thumb wheel was turned too quickly. To resolve this, clean or disinfect the suction system. If cleaning doesn't suffice, replace the suction system.
  • D
    Debbie CastroAug 19, 2025
    How to fix ejector key if it does not work on BRAND micro Laboratory Equipment?
    • K
      Kimberly ZhangAug 19, 2025
      If the ejector key on your BRAND Laboratory Equipment isn't working, it could be improperly positioned. To fix this, remove the clip, hold the instrument loosely, and quickly press down the ejector key. Then, slide the clip back on.
  • A
    Angela BradyAug 22, 2025
    What to do if pipette is ejected from the adapter of BRAND micro Laboratory Equipment?
    • F
      fcollinsAug 22, 2025
      If the pipette is being ejected from the adapter of your BRAND Laboratory Equipment, the adapter may be wet or greasy. Try cleaning or disinfecting the suction system and ensure it's completely dry. If the issue persists, you may need to replace the suction system.
  • T
    Tamara HickmanAug 25, 2025
    How to remove broken pipette in the adapter of BRAND micro?
    • B
      bonillaedwardAug 26, 2025
      If you have a broken pipette in the adapter of your BRAND Laboratory Equipment, it's likely because the pipette was damaged or forced into the adapter. Carefully remove the glass splinters using forceps. If necessary, replace the adapter.
  • S
    Sharon AdkinsAug 28, 2025
    What to do if there is liquid in the suction system of BRAND Laboratory Equipment?
    • K
      kpooleAug 28, 2025
      If liquid gets into the suction system of your BRAND Laboratory Equipment due to turning the adjustment wheel too quickly, you should clean or disinfect the suction system. If that doesn't resolve the issue, consider replacing it.
  • G
    Gabrielle CalhounAug 30, 2025
    Why is the suction system of BRAND micro not working or pipette dripping?
    • D
      Dennis HayesAug 31, 2025
      If the suction system of your BRAND Laboratory Equipment is damaged, causing it to not work or the pipette to drip, the suction system should be replaced.
  • R
    Rachel CarsonSep 3, 2025
    How to remove a broken pipette from the adapter in BRAND micro?
    • H
      Howard HillSep 3, 2025
      If a pipette breaks off in the adapter of your BRAND Laboratory Equipment because it was damaged or pushed in with too much force, remove the glass fragments with tweezers. If necessary, replace the adapter.
  • S
    Shelby RosalesSep 6, 2025
    Why BRAND Laboratory Equipment does not draw in liquid or the pipette drips?
    • K
      Kathleen StaffordSep 6, 2025
      If your BRAND Laboratory Equipment isn't drawing in liquid or the pipette drips, it might be due to a damaged suction system. In this case, you should replace the suction system.
  • B
    Brittany ThomasSep 8, 2025
    What to do if the ejector button does not work on BRAND Laboratory Equipment?
    • N
      nnielsenSep 9, 2025
      If the ejector button on your BRAND Laboratory Equipment isn't working because it's not positioned properly, remove the clip. Hold the instrument loosely and briefly press the ejector button, then slide the clip back on.
  • M
    malloryrobertsSep 12, 2025
    What to do if the pipette is ejected from the adapter of BRAND Laboratory Equipment?
    • C
      Candice JonesSep 12, 2025
      If the pipette is ejected from the adapter of your BRAND Laboratory Equipment because the adapter is too moist or greasy, clean or disinfect the suction system and let it dry. If the issue persists, replace the adapter.

Summary

Safety Instructions

Function and Application

Pipetting Procedure

Attaching the Pipette

Instructions on how to securely fit pipettes to the controller without damage.

Adjusting Suction Volume

Guide to setting the desired volume using the thumb wheel for different pipette sizes.

Drawing Up Liquid

Step-by-step process for aspirating liquids, emphasizing contamination prevention.

Discharging Liquid

Procedures for dispensing liquids accurately, including specific methods for 'In' calibrated pipettes.

After Pipetting

Cleaning and Autoclaving

Repair and Servicing

Return for Repair

Troubleshooting