AESCULAP® 3D EinsteinVision® 3.0
Flow 50 Insufflator
On start-up - First screen offers surgical specialties
Select appropriate Indication- followed by ”Veress” or “High Flow” to move to main screen.
(Cardiac mode has 2 functions - “Vessel Harvesting” or “Field Flooding” (Field flooding is flow only)
Pre-set pressure can be set up to 30 mmHg
Nb. 15 mmHg is set as the Safety Threshold “Safety Threshold” will flash repeatedly
Press + when the safety threshold message has gone off, to take pressure beyond 15 mmHg
Press and hold the Flow + or buttons to go between “Low Flow”, Medium Flow” and “High Flow”
Low Flow = 3 L/min Medium Flow = 20 L/min High Flow = 40 L/min
Insufflation is started and stopped using the “Start” button (Bottom Right)
(When insufflation is active, button displays “Stop”. Insufflation is stopped by pressing “Stop”
CO
2
volume delivered is reset by pressing the reset button
Optional Smoke
Evacuation Filter
Housing
Active Smoke Evacuation
Actual
Actual
Volume
& Reset
System Settings
Insufflation
Gas Outlet
Heated Gas
Element
(Optional)
Start / Stop
Insufflation
Smoke Evacuation
Footswitch
(Optional)
Pressure & Flow
Pre-sets
On / Off