Doc. HD40-0007-02 En-Eu L-4
Cleaning
Regular cleaning should be performed in accordance with the instructions below.
WARNING: This device can expose you to electrocution. Do not immerse
the Z1, Power Adapter, Power Cord or Battery in water. Always unplug these
devices before cleaning (only as instructed in this User Guide) and make sure
they are dry before plugging back in.
CAUTION: Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or antibacterial soaps or scented oils to clean the Z1.
Never attempt to clean the Power Adapter, PowerShell, Battery or accessories
with any of these solutions. These solutions may create a shock hazard, cause
damage and reduce the life of these products.
Daily
1. Remove the CPAP Tube and the mask from the Z1.
2. Refer to the instructions for cleaning, maintenance and replacement provided with
your CPAP Tube and mask.
3. Before next use, reconnect the CPAP Tube and the Tube Adapter to the Z1 and mask.
As Needed
1. Clean the exterior of the Z1 with a soft damp cloth and mild detergent and wipe
clean. Never submerge the Z1 in water. Do not allow soap or water to enter the Z1.
Filter Replacement
The Filter should be inspected weekly for holes and blockage by dirt and dust, and should be
replaced every month (or more often if necessary). The Filter is accessed by removing the End Cap
as described in Assembling and Powering Up: Filter and End Cap”.
NOTE: The Z1 uses a specially designed Filter that is only available from
Breas. To purchase replacement lters visit www.breas.com.
WARNING: Do not attempt to use any air lter which is not supplied by
Breas for the Z1. The Z1 is designed to only work with approved materials. A
lter which is not approved may cause damage to the Z1 from overheating.
CAUTION: Do not operate the Z1 without the Filter. The Z1 must
operate with a lter installed; failure to do so may damage the Z1 and allow
impurities into the respiratory system.
CPAP Tube Replacement
The CPAP Tube should be replaced according to the instructions provided with the CPAP Tube.