28
Doc. HD40-0007-02 En-Eu L-4
Technical Specications
30W Power Adapter Unit Input range 100–240V, 50–60Hz
Typical power consumption 20W (30VA)
Maximum power consumption 35W (70VA)
Operating temperature 5
º
C to 35
º
C (41
º
F to 95
º
F)
Operating humidity 10–80% Relative Humidity, Non-Condensing
Operating altitude Sea level to 2400m (8000ft)
Operating atmospheric pressure 101 to 75.3 kPa
Storage and transport temperature -20
º
C to +60
º
C (-4
º
F to +140
º
F)
Storage and transport humidity 10–80% Relative Humidity, Non-Condensing
Storage and transport altitude No limit
Storage and transport atm. pressure No limit
Electromagnetic compatibility Compatible with IEC 60601-1-2
Aircraft use Complies with RTCA DO-160G Standard for use in
ight
IEC 60601-1 classication Class II (double insulation), Type BF
Available modes Standby, CPAP, Auto-CPAP and CPAP/Auto-CPAP
with Ramp
Sound pressure level < 26 dBA
Nominal dimensions (L x W x H) 16.46 cm x 8.38 cm x 5.13 cm (6.48” x 3.30” x 2.02”)
Weight 284 g (10 oz.)
Housing construction Flame retardant engineered thermoplastic
Supplemental oxygen Not for use with oxygen
Standard air lter Polyester
Air outlet The 22 mm air outlet complies with ISO 5356-1
PowerShell Battery Specications 45 Watt-hours capacity
Lithium Ion Battery Pack
4S1P
Weight approx. 225 g (8 oz.)
Expected Service Life Z1, PowerShell and Power Supply: 5years
Tube Adapter, USB Cable and End Cap: 2years
Filter: See Care Instructions
CPAP Tube: See CPAP Tube Instructions
Performance Specications
Operating pressure range 4 to 20 cm H
2
O
Maximum single fault steady pressure state 30 cm H
2
O
Deliverable pressure accuracy ±0.6 cm H
2