27
Doc. HD40-0007-02 En-Eu L-4
Recommended separation distances between portable and mobile RF communications
equipment and the Z1 CPAP System
The Z1 CPAP System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Z1 CPAP System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Z1 CPAP System as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d=1,2√P
80 MHz to 800 MHz
d=1,2√P
800 MHz to 2,5 GHz
d=2,3√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in metres (m) can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected
by absorption and reection from structures, objects and people.
WARNING: Please note that mobile RF communications equipment can
eect medical electrical equipment.
Wireless Quality of Service for Z1 Bluetooth Communication with Z1 Mobile App
The Z1 and Mobile App utilize only a very small fraction of the available bandwidth and through-
put. Based on this, the Z1 and Mobile App are highly tolerant of reductions in wireless quality of
service and performance will not be aected until service is essentially lost; Wireless connectivity
is not part of the essential performance of the Z1 performance. All functions of the Z1 can be
performed on the Z1 unit itself; The Mobile App and Z1 are tested for coexistence with other in-
band radio frequency sources; The Bluetooth hardware module is FCC listed, and complies with
FCCPART15 SUBPARTC.