26
Doc. HD40-0007-02 En-Eu L-4
Guidance and manufacturer’s declaration — electromagnetic immunity
The Z1 CPAP System is intended for use in the electromagnetic environment specied below. The
customer or the user of the Z1 CPAP System should assure that it is used in such an environment.
Immunity test IEC60601 test
level
Compliance
level
Electromagnetic environment,
guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the Z1 CPAP
System, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC61000-4-6
3 Vrms
150kHz to 80MHz
3 Vrms
150kHz to
80MHz
d=1,2√P
Radiated RF
IEC61000-4-3
3 V/m
80MHz to 2,5GHz
3 V/m
80MHz to
2,5GHz
d=1,2√P 80MHz to 800MHz
d=2,3√P 80MHz to 2,5GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from xed RF transmit-
ters, as deter-mined by an electromag-
netic site survey,
a
should be less than
the compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected
by absorption and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to xed RF transmitters, an electromagnetic site survey should be considered. If the measured
eld strength in the location in which the Z1 CPAP System is used exceeds the applicable RF
compliance level above, the Z1 CPAP System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Z1 CPAP System.
b
Over the frequency range 150kHz to 80MHz, eld strengths should be less than 3V/m.