25
Doc. HD40-0007-02 En-Eu L-4
Guidance and manufacturer’s declaration — electromagnetic immunity
The Z1 CPAP System is intended for use in the electromagnetic environment specied below. The
customer or the user of the Z1 CPAP System should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic envi-
ronment - guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power
supply lines
±1 kV for input/out-
put lines
±2 kV for power
supply lines
±1 kV for input/out-
put lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC61000-4-11
<5 % U
T
(>95 % dip in U
T
) for
0,5cycle
40% U
T
(60% dip in U
T
) for
5cycles
70% U
T
(30% dip in U
T
) for
25cycles
<5% U
T
(>95 % dip in U
T
) for
5sec
<5 % U
T
(>95 % dip in U
T
) for
0,5cycle
40% U
T
(60% dip in U
T
) for
5cycles
70% U
T
(30% dip in U
T
) for
25cycles
<5% U
T
(>95 % dip in U
T
) for
5sec
Mains power quality
should be that of a typical
commercial or hospital
environment, If the user
of the Z1 CPAP System
requires continued oper-
ation during power mains
interruptions, it is recom-
mended that the Z1 CPAP
System be powered from
an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
eld
IEC61000-4-8
3A/m 3A/m Power frequency mag-
netic elds should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.