EasyManua.ls Logo

Bristol 671 - Page 34

Bristol 671
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 671 User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 34
*ESR?
The *ESR (event status register) query returns a value which encodes
the bits in the event status register. If any bits are set in the ESR, then
the ESR summary bit will be set in the STB.
Returns an integer which is the sum of all the bit values for those bits
that are set. See Event Status Register table below.
Event Status Register
BIT
BIT VALUE
CONDITION
7
128
Power On (PON)
6
64
Not Used
5
32
Command Error (CME)
4
16
Execution Error (EXE)
3
8
Device Dependent Error (DDE)
2
4
Query Error (QYE)
1
2
Not Used
0
1
Operation Complete (OPC)
*IDN?
The *IDN (identification number) query returns a string value which
contains the instrument type, serial number, and firmware version. The
third value is the instrument serial number. The last value is the
imbedded software version.
BRISTOL WAVELENGTH METER, 671, 1135, 1.03

Related product manuals