Rev:Y2,August,2020 ATP_ASP Price Computing Scales Operation Manual
- 10 -
6. CALIBRATION
1 Turn on the scale.
2 Press and hold key during the self test.
3 Display will show Pn briefly PIN is activated.
4 Press the PIN number to get into the parameter menu.The default PIN number is
0000
5 Press numeric key 0 four times , display will be shown Pn - - - -
6 Press key to confirm, It will enter into the parameter settings and will be
Show MENU UNIT KG
7 If calibrating in kg.,continue step 9
8 If calibrating in lb.,press key ,lb will be show on the display
9 Press key to confirm , display will be show F0 CAL
Then press the calibration switch to enter calibration,switch is locating below the
scale.
10 Press key to confirm for calibration, display will be show UNLOAD
11 Remove the goods from the platform and make sure STABLE indication has been
enable.
12 Press key to confirm,Display will be show ,CAP LOAD XX
13 If necessary select the calibration weight value from the test weights list by
pressing key and place the calibration weight on the platform.
14 Press key to confirm when display gets stable,display will be show PASS
15 Finished calibration and display will come to F0 CAL
16 Remove calibration weight from the platform.
17 Press key to go back to the weighing mode.