Brookeld Engineering Laboratories, Inc. Page 12 Manual No. M00-151-I0614
HA Series Viscometer
Spindle Number
* 1 2 3 4 5 6 7
0.5
1
2
2.5
4
5
10
20
50
100
400
200
100
80
50
40
20
10
4
2
0.5
1
2
2.5
4
5
10
20
50
100
1.6K
800
400
320
200
160
80
40
16
8
0.5
1
2
2.5
4
5
10
20
50
100
4K
2K
1K
800
500
400
200
100
40
20
0.5
1
2
2.5
4
5
10
20
50
100
8K
4K
2K
1.6K
1K
800
400
200
80
40
0.5
1
2
2.5
4
5
10
20
50
100
16K
8K
4K
3.2K
2K
1.6K
800
400
160
80
0.5
1
2
2.5
4
5
10
20
50
100
40K
20K
10K
8K
5K
4K
2K
1K
400
200
0.5
1
2
2.5
4
5
10
20
50
100
160K
80K
40K
32K
20K
16K
8K
4K
1.6K
800
HB Series Viscometer
Spindle Number
* 1 2 3 4 5 6 7
0.5
1
2
2.5
4
5
10
20
50
100
1.6K
800
400
320
200
160
80
40
16
8
0.5
1
2
2.5
4
5
10
20
50
100
6.4K
3.2K
1.6K
1.28K
800
640
320
160
64
32
0.5
1
2
2.5
4
5
10
20
50
100
16K
8K
4K
3.2K
2K
1.6K
800
400
160
80
0.5
1
2
2.5
4
5
10
20
50
100
32K
16K
8K
6.4K
4K
3.2K
1.63K
800
320
160
0.5
1
2
2.5
4
5
10
20
50
100
64K
32K
16K
12.8K
8K
6.4K
3.2K
1.6K
640
320
0.5
1
2
2.5
4
5
10
20
50
100
160K
80K
40K
32K
20K
16K
8K
4K
1.6K
800
0.5
1
2
2.5
4
5
10
20
50
100
640K
320K
160K
128K
80K
64K
32K
16K
6.4K
3.2K
=Spindle =SpindleSpeed
=Factor
K = 1000*Optional
II.4 Consideration for Making Measurements
In taking viscosity measurements with the Dial Viscometer, there are two considerations which
pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range from
10% to 100% for any combination of spindle/speed rotation.
2) Viscosity measurements should be taken under laminar ow conditions, not under turbulent
ow conditions.
The rst consideration has to do with the accuracy of the instrument. All Dial Viscometers have
a full scale range precision of ± 1% for any spindle/speed combination. We discourage taking
readings below 10% of range because the potential viscosity error of ± 1% is a relatively high
number compared to the instrument reading.