Brookeld Engineering Laboratories, Inc. Page 6 Manual No. M00-151-I0614
Please check to be sure that you have received all components, and that there is no damage. If you are
missing any parts, please notify Brookeld Engineering or your local Brookeld agent immediately.
Any shipping damage must be reported to the carrier.
I.2 Utilities
VAC; Hz Limits 100 - 240 VAC; 50/60 Hz ± 5%
Power Supply 15 Watts, Class II certied plug-in power supply rated:
12V @ 1.25A
Main supply voltage uctuations are not to exceed 10% of the nominal supply voltage.
I.3 Specications
Speeds:
LVT 60, 30, 12, 6, 3, 1.5, 0.6, 0.3
RVT 100, 50, 20, 10, 5, 4, 2.5, 2, 1, 0.5
HAT 100, 50, 20, 10, 5, 4, 2.5, 2, 1, 0.5
HBT 100, 50, 20, 10, 5, 4, 2.5, 2, 1, 0.5
Accuracy: ± 1% of Full Scale Range in use
(See Appendix C for details)
Repeatability: ± 0.2% of Full Scale Range in use
Weight:
Gross Weight 20 lb 9 kg
Net Weight 17 lb 7.7 kg
Carton Volume 1.65 cu 0.05 m
3
Operating Temperature Range: 5°C to 40°C (41°F to 104°F)
Operating Relative Humidity: 80% up to 30°C decreasing linearly to 50% at 40°C
Electrical Certications:
Conforms to CE Standards:
BSEN 61326: Electrical equipment for measurement, control and laboratory use - EMC
requirements
BSEN 61010-1: Safety requirements for electrical equipment, for measurement, control
and laboratory use
Notice to customers:
is symbol indicates that this product is to be recycled at an appropriate collection center.