Brookeld Engineering Laboratories, Inc. Page 19 Manual No. M00-151-I0614
Brookeld Viscosity Standard Fluid General Information
We recommend that Brookeld Viscosity Standard Fluids be replaced on an annual basis, one year from
date of initial use. Exposure to outside contaminants, such as solvent, standard of dierent viscosity
or other foreign material, requires replacement annually. Oil Fluids do have an expiration date on the
label that should be adhered to.
Viscosity Standard Fluids may be stored under normal laboratory conditions. Disposal should be in
accordance with state, local and federal regulations as specied on the material safety data sheet.
Brookeld Engineering Laboratories does not recertify Viscosity Standard Fluids. We will issue duplicate
copies of the Certicate of Calibration for any uid within two years of the purchase date.
Brookeld Viscosity Standard Fluids are reusable provided they are not contaminated. Normal practice
for usage in a 600 mL beaker is to return the material from the beaker back into the bottle. When using
smaller volumes in accessories such as Small Sample Adapter, UL Adapter or ermosel, the uid is
normally discarded.
e following sections explain how to perform a calibration check on your viscometer using dierent
spindle geometries. For additional assistance, videos on these procedures are available on the Brookeld
website, www.brookeldengineering.com.
Calibration Procedure for LV #1-3(61-63) and RV,HA,HB #1-6(01-06) Brookeld Spindles
Please note that the LV #4 and the RV, HA and HB #7 spindles have been omitted from this procedure.
Brookeld does not recommend the use of these spindles to perform a calibration check on your
instrument. Reasons pertain to the small amount of spindle surface area that makes contact with
the viscosity standard. e diculty establishing the immersion mark precisely and the need for
precise temperature control at 25˚C in the immediate vicinity of the spindle.
Follow these steps by using one of the recommended spindles to verify calibration of your instrument:
1) Place the viscosity standard uid (in a 600 mL low form beaker) into the water bath.
2) Lower the Viscometer into measurement position (with guard leg if LV or RV series Viscometer
is used).
3) Attach the spindle to the Viscometer. If you are using a disk-shaped spindle, avoid trapping air
bubbles beneath the disk by rst immersing the spindle at an angle, and then connecting it to
the Viscometer.
4) e viscosity standard uid, together with the spindle and guardleg (if used), should be immersed
in the bath for a minimum of 1 hour, stirring the uid periodically, prior to taking measurements.
5) Aer 1 hour, check the temperature of the viscosity standard uid with an accurate thermometer.
6) If the uid is at test temperature (±0.1°C of the specied temperature, normally 25°C), measure
the viscosity and record the Viscometer reading.
Note: The spindle must rotate at least ve (5) times before readings are taken.