Icon System Manual T02985_02_03 15
MRI System Overview
2.6 Animal Positioning System
The animal support for the ICON is modular and consists of the basic animal holder
(see section 2.6.1) and exchangeable front pieces; the actual animal beds
(see section 2.6.2). Every animal bed has a RF Coil included, supporting various
applications:
mouse body (length: 65 mm)
mouse whole body (length: 95 mm)
mouse head
rat body
60 mm rat body (magnet mounted)
rat head
The magnet and gradient coil iso-center position is reached when the ICON animal
holder is positioned at the zero position on the magnet front flange.
2.6.1 Animal Holder
The animal holder has permanently installed animal anesthesia, care and supervision
lines and tubes.
Features of the animal holder:
Reproducible animal positioning in the magnet (adjustable position and rotation)
Ergonomic handle at the center of gravity
Fixation in the magnet with a knurled screw
Connection plate for the animal anesthesia, care and supervision lines and tubes
Connection of the RF Coil
Tuning connection for the RF Coil
shifting rod for moving the body coils once mounted in the magnet
RF shielding functionality with filter plate in the rear connection panel
(not visible, under the black hood)