Icon System Manual T02985_02_03 33
Safety (English)
3.7 RF Coils and Accessories
Under normal operation conditions no risk to the operator arises from the usage of Bruker
tested RF Coils and accessories. Nevertheless, the following safety aspects need to be
considered.
WARNING
Risk of injury or death.
Do not use RF Coils with damaged cables or housing.
Do not use RF Coils outside the magnet when scanning.
CAUTION
Risk of personal injury or injury to the animal.
Leaking water may lead to arcing or electric shocks in the RF Coil or animal bed.
Before preparing the animal or sample always make sure that the warm water tubes
are not leaking. In particular, visually inspect if water entered the animal bed or RF
Coil housing. If water was entering the animal bed or RF Coil,
swab the water completely with a dry and soft cloth;
any liquids in RF Coils must be completely evaporated at temperatures 40°C
before using it.
RF Coils and accessories that are not tested and released by Bruker for compatibility with
the MR system bear a potential risk.
CAUTION
Risk of injury.
Do not use RF Coils or accessories that are not released by Bruker.