Check the level of the filling solution. It should be about 1cm below the filling
orifice. To add solution, see section 3.3.
The probe should be stored standing upright. The electrode tip should be immersed
in a solution of 3-Molar KCL or a Buffer Solution between pH 4 and pH 7. The two
rubber "T" stoppers should be inserted. At no time should the electrode be allowed
to rest on the tip.
NOTE: The rubber T-stoppers should be left in place during autoclaving,
but removed during regular usage.
Certain types of Mycelia can grow on the liquid junction. These growths can be
prevented from forming by the addition of a few drops of Formaldehyde (up to 1%)
to the 3-Molar KCL solution.
4.6 THE AGITATOR BEARING HOUSING (Fig. 23)
Every 3-6 months, the ball bearings and the shaft seals in the bearing housing should be
checked and cleaned. Replace the worn out bearings and shaft seals.
4.7 REPLACEMENT PART LIST
Vessel (1.25L) Assembly M1169-2000
Vessel (2.5L) Assembly M1169-2001
Vessel (5L) Assembly M1169-2003
Inlet Filter P0200-0491 (37mm Disc)
Exhaust Filter P0200-0490 (50mm Disc)
RTD Assembly M1169-8002
pH Probe (1.25L, 2.5L) P0720-5325
pH Probe Cable (1.25L, 2.5L) P0720-2095
pH Probe (5L) P0720-5021
pH Probe Cable P0720-2095
pH Electrode K9 Cap P0720-5317
pH Probe Electrolyte P0860-0130
DO Probe (1.25L) (Phoenix) P0720-5440
DO Probe (2.5L) (Phoenix) P0720-5450
DO Probe (5L) (Phoenix) P0720-5460
DO Probe Cable (Phoenix) P0720-2202
DO Probe (1.25L) (Ingold) P0720-5560
DO Probe (2.5L) (Ingold) P0720-5561
DO Probe (5L) (Ingold) P0720-5561
DO Cable (Ingold) P0720-2202
DO Service Kit, Ingold P0720-5569
DO Electrode Cap, Ingold P0720-5567
Teflon Washer for Ingold DO Probe P0100-9780
O-Ring for Ingold DO Probe P0280-9163
Seal Washer for Ingold DO Probe M1016-0890
Antifoam Probe F5-137
Hose Connect Plug M1151-9518
Hose Connect Open M1151-9519
Seal, Vessel Side Port M1151-9419
Nut, Vessel Side Port M1151-9420
Fuse, 10 Amp, 115V EF-143
Fuse, 6 Amp, 220V P0380-3140
Glass Vessel 1 1/4L M1151-9904
Glass Vessel 2 1/2L M1152-9903